Keep the original entry and read its definition
Copy the exact test or analyte name and result text from the laboratory report, including any less-than sign, qualifier and footnote. Ask the issuing laboratory to explain an undefined entry.
- For ND or not detected, request the report's definition and the applicable numerical threshold. Do not assume every laboratory uses ND to mean below LOD or below LOQ.
- Keep a missing or unreported test separate from a non-detection. Ask whether the requested item was tested and obtain its report entry.
Separate detection from quantification
LOD, the limit of detection, describes detection capability. LOQ, the limit of quantitation or quantification, concerns the ability to give a quantitative result with acceptable precision. These address different questions.
- An entry of <LOQ may include an analyte that was detected but could not be quantified at the required performance level. Ask the laboratory what the entry establishes for this result; it does not by itself prove absence.
- Request the applicable LOD, LOQ and any separate reporting limit for each requested analyte, with units. Ask whether these describe the complete analytical method and actual reported result, rather than only an instrument's detection capability.
Give the quality lead the limits and reporting conditions
Keep the material, sample and lot link, method reference and version, unit, reporting basis and laboratory notes together. Where as-received or dry-matter reporting is used, identify it explicitly for both the result and the applicable limit.
- Ask the laboratory to clarify the limits applicable to this material and result, including any qualifications stated in the report. Request an explanation of any unit or basis conversion needed for the review.
- Give your quality lead the agreed purchasing criterion, original result and applicable limits. Ask whether the report's qualifications and the applicable decision rule support the intended purchasing decision, and what clarification or further evidence is needed.
Send one clarification row per test item
For organic soy expeller cake, soybean meal or forage ingredients, send our MUSE team the complete report and unresolved items through the contact options below. We can organise supplier questions and requests for the issuing laboratory's explanation.
- Use this row: ingredient and lot | report and sample identifiers | test or analyte | original result text | unit and basis | method and version | applicable LOD, LOQ and reporting limit | report definition and notes | question and responsible person.
- Keep the original report beside written clarifications. Record the quality lead's purchasing decision separately; retain unresolved questions without filling them with a zero or an assumed acceptance.
Buyer questions
Can I put zero in the comparison sheet when the COA says ND?
Keep ND and its report-defined meaning, threshold, unit and basis. It is not a measured zero. If a calculation needs a numerical input, ask your quality lead how that qualified result should be handled and preserve the original entry.
Do two suppliers reporting ND provide equivalent evidence?
Not necessarily. Compare the exact analyte, method, material, applicable limits, units, reporting basis and report definitions. Two identical abbreviations can represent different reporting conditions.